SmartChem 200
Overview
Article Owner: Scott Gallagher
Purpose: Document SmartChem 200 software setup, configuration, and support procedures
Audience: IT support staff, lab administrators, and faculty using SmartChem 200 in BI 339
Software Overview
This section provides information about the software.
Software Name: SmartChem 200
Latest Request: Reinstallation and validation of licensing and user access on David Shull’s laptop (BI 339)
Version: 3.1.65
Import Date: June 12, 2025
End-of-Life Date: Unknown
Purpose: Nutrient analysis and data collection via chemical analyzer hardware
Owning Department: Biology and ESCI
Software Administrator: Scott Gallagher
Subject Matter Experts: David Shull and David Hooper
Vendor contact: service@kpmanalytics.com | +1 (774) 399‑0500 (Press 2, then 2 for tech support)
Software Scope: Local Install on CENV-M-8CFCSG3
OS Type: Windows 11
CPU Architecture Support: x64
Software Management
Licensing Overview
License Type: Hardware USB Dongle
License Quantity: Single-User, tied to the dongle
License Expiration: Perpetual
License / Account Creation
Software runs via physical dongle; no user accounts required.
Default credentials for support reference:
Username: Westco
Password: joe
Deprovisioning
Remove software via Control Panel and securely store dongle in lab IT drawer or return to Scott Gallagher for inventory.
Uninstall Command: haspdinst.exe -fr -kp -purge -fss
Files | Description |
---|---|
| Key for USB Dongle |
Supported version |
Windows Deployment Details
Install Command: Run Executable
Uninstall Command: Remove via appwiz.cpl
Detection Method: Check for Hardlock service and presence of C:\SmartChemNew\DB\easychem
Does new version uninstall old version?
No
User Experience
Installation Behavior: Silent after launch; CLI-based driver installer
Logon Requirement: Administrator required
Installation Visibility: Partial (console visible during haspdinst execution)
Estimated Installation Time (min): 5
Prerequisites / Dependencies
Disable Memory Integrity (Core Isolation) via Group Policy or Registry
Hardlock dongle must be removed during driver installation
Lab Use Configuration
Date Format Required: dd/mm/yyyy
Power Settings:
Display and Sleep timers disabled while on AC
SmartChem Configuration:
Enable Mo3 only
Uncheck “Do not show concentrated results”
User Account Notes
Any user with access to the system and SmartChem install folder should be able to view runs.
All critical data resides in the shared DB path (C:\SmartChemNew\DB)
and is not tied to individual profiles
Backup and Maintenance
Scheduled Task: To be coordinated with David Shull for automated backups of DB files to designated storage
Recommend weekly export or sync to Ceph/OneDrive for redundancyPatching
Patching
No known automated patching mechanism. Contact vendor for updates.
Troubleshooting
Common Issues
Driver install hangs at "Please wait"
Hardlock dongle not recognized
Access issues across user accounts
Error Messages
Error 53: Missing
HLVDD.DLL
Error 193: Application architecture mismatch
Status Code 48 5 774: Hardlock service failure
Known Workarounds
Disable Memory Integrity via Group Policy
Fully uninstall and purge driver before reinstalling
Copy DB files (
easychem
,easystore
,recalc
) from backup as needed
Escalation Path
onfirm local config (driver install, dongle recognition, permissions)
Contact Scott Gallagher